dc.identifier.citationMarkarian, R. A., Vasconcelos, E., Kim, J. H., Attard, N. J. & Cortes, A. R. G. (2022). Effect of different milling devices on marginal fit of CAD/CAM zirconia copings on implant stock abutments. International Journal of Prosthodontics, 35(4), 420-424.en_GB
dc.identifier.urihttps://www.um.edu.mt/library/oar/handle/123456789/102840-
dc.description.abstractPurpose: To compare the marginal fit and internal surface roughness of CAD/CAM zirconia copings milled with three- and five-axis milling devices. Materials and Methods: A total of 40 titanium implant stock abutments (4.8 mm in diameter, 4 mm in height) screwed to dental implants (4.1 mm in diameter) embedded in resin were considered phantoms and included in this in vitro study. All 40 phantoms were scanned with the same intraoral scanner, and virtual wax-ups of zirconia copings were created from these images and exported as STL files. Two copings were milled from each resulting STL file—one using a three-axis milling device and the other using a five-axis milling device. After milling, zirconia copings underwent high-speed sintering before being analyzed for marginal fit (ie, marginal gap measurement) and observation of internal surface roughness with scanning electron microscopy (SEM). Statistical comparisons between groups were assessed with Mann-Whitney U test. Results: Median marginal gap values were 34.80 μm (95% CI: 0.00 to 173.98) for the five-axis milling device group and 141.97 μm (95% CI: 82.13 to 163.46) for the three-axis milling device group. A statistically significant difference in marginal gap was found between both milling device groups (P = .039). In addition, qualitative SEM analysis indicated a higher internal surface roughness for the three-axis milling device group. Conclusion: Within the limitations of this study, the present findings suggest that five-axis milling devices outperform three-axis milling devices for milling CAD/CAM zirconia copings from intraoral scans of implant stock abutments.en_GB
dc.language.isoenen_GB
dc.publisherQuintessence Publishing Co., Inc.en_GB
dc.rightsinfo:eu-repo/semantics/restrictedAccessen_GB
dc.subjectDental implantsen_GB
dc.subjectComputer-aided designen_GB
dc.subjectDental abutmentsen_GB
dc.subjectDental technologyen_GB
dc.titleEffect of different milling devices on marginal fit of CAD/CAM zirconia copings on implant stock abutmentsen_GB
dc.typearticleen_GB
dc.rights.holderThe copyright of this work belongs to the author(s)/publisher. The rights of this work are as defined by the appropriate Copyright Legislation or as modified by any successive legislation. Users may access this work and can make use of the information contained in accordance with the Copyright Legislation provided that the author must be properly acknowledged. Further distribution or reproduction in any format is prohibited without the prior permission of the copyright holder.en_GB
dc.description.reviewedpeer-revieweden_GB
dc.identifier.doi10.11607/ijp.7069-
dc.publication.titleInternational Journal of Prosthodonticsen_GB
Appears in Collections:Scholarly Works - FacDenCDHO
Scholarly Works - FacDenDS



Items in OAR@UM are protected by copyright, with all rights reserved, unless otherwise indicated.