Please use this identifier to cite or link to this item: https://www.um.edu.mt/library/oar/handle/123456789/103388
Full metadata record
DC FieldValueLanguage
dc.contributor.authorGatt, Alexander-
dc.contributor.authorvan Veen, Joost J.-
dc.contributor.authorWoolley, Anita M.-
dc.contributor.authorKitchen, Steve-
dc.contributor.authorCooper, Peter-
dc.contributor.authorMakris, Michael-
dc.date.accessioned2022-11-04T07:27:26Z-
dc.date.available2022-11-04T07:27:26Z-
dc.date.issued2008-
dc.identifier.citationGatt, A., van Veen, J. J., Woolley, A. M., Kitchen, S., Cooper, P., & Makris, M. (2008). Thrombin generation assays are superior to traditional tests in assessing anticoagulation reversal in vitro. Thrombosis and Haemostasis, 100(08), 350-355.en_GB
dc.identifier.urihttps://www.um.edu.mt/library/oar/handle/123456789/103388-
dc.description.abstractEven though new anticoagulants are being devised with the notion that they do not require regular monitoring, when bleeding occurs, it is important to have an antidote and a reliable test to confirm whether the anticoagulant effects are persisting. We examined the effects of five heparinoids, unfractionated heparin (UFH), tinzaparin, enoxaparin, danaparoid and fondaparinux on the traditional APTT and anti-Xa assays as well as on the calibrated automated thrombogram (CAT). We also studied the ability of protamine sulphate (PS), NovoSeven®, FEIBA® and FFP to reverse maximum anticoagulation induced by the different heparinoids.The CAT was the only test to detect the coagulopathy of all the anticoagulants.PS produced complete reversal of UFH, and this could be monitored with all three tests. Tinzaparin can also be completely neutralised in vitro with high doses of PS, but the maximum enoxaparin reversal achieved with PS is only approximately 60%. Fondaparinux does not significantly affect the APTT and PS has no significant effect on its reversal. Only NovoSeven was able to correct the fondaparinux induced CAT abnormalities whilst having no effect on the anti-Xa level. None of the reversal agents was very effective in danaparoid spiked plasma but NovoSeven,at high dose,increased the ETP by 40% and reduced the anti-Xa level from 0.93 to 0.78 IU/ml.We conclude that the CAT is superior to the traditional coagulation tests in that it not only detects the coagulopathy of all the heparinoids but can be also be used to monitor their reversal.en_GB
dc.language.isoenen_GB
dc.publisherGeorg Thieme Verlagen_GB
dc.rightsinfo:eu-repo/semantics/restrictedAccessen_GB
dc.subjectThrombinen_GB
dc.subjectAnticoagulants (Medicine) -- Testingen_GB
dc.subjectAnticoagulant reversal agents -- Testingen_GB
dc.subjectHeparin -- Therapeutic useen_GB
dc.subjectBlood coagulation factorsen_GB
dc.titleThrombin generation assays are superior to traditional tests in assessing anticoagulation reversal in vitroen_GB
dc.typearticleen_GB
dc.rights.holderThe copyright of this work belongs to the author(s)/publisher. The rights of this work are as defined by the appropriate Copyright Legislation or as modified by any successive legislation. Users may access this work and can make use of the information contained in accordance with the Copyright Legislation provided that the author must be properly acknowledged. Further distribution or reproduction in any format is prohibited without the prior permission of the copyright holder.en_GB
dc.description.reviewedpeer-revieweden_GB
dc.identifier.doi10.1160/TH07-05-0357-
dc.publication.titleThrombosis and Haemostasisen_GB
Appears in Collections:Scholarly Works - FacM&SPat



Items in OAR@UM are protected by copyright, with all rights reserved, unless otherwise indicated.