Please use this identifier to cite or link to this item: https://www.um.edu.mt/library/oar/handle/123456789/104431
Title: Validity and reliability of a new ankle dorsiflexion measurement device
Authors: Gatt, Alfred
Chockalingam, Nachiappan
Keywords: Joints -- Range of motion -- Measurement
Ankle -- Movement disorders -- Diagnosis
Foot -- Abnormalities
Joints -- Physiology
Goniometers
Issue Date: 2013
Publisher: Sage Publications Ltd.
Citation: Gatt, A., & Chockalingam, N. (2013). Validity and reliability of a new ankle dorsiflexion measurement device. Prosthetics and orthotics international, 37(4), 289-297.
Abstract: Background: The assessment of the maximum ankle dorsiflexion angle is an important clinical examination procedure. Evidence shows that the traditional goniometer is highly unreliable, and various designs of goniometers to measure the maximum ankle dorsiflexion angle rely on the application of a known force to obtain reliable results. Hence, an innovative ankle dorsiflexion measurement device was designed to make this measurement more reliable by holding the foot in a selected posture without the application of a known moment.
Objectives: To report on the comprehensive validity and reliability testing carried out on the new device.
Methods: Following validity testing, four different trials to test reliability of the ankle dorsiflexion measurement device were performed. These trials included inter-rater and intra-rater testings with a controlled moment, intra-rater reliability testing with knees flexed and extended without a controlled moment, intra-rater testing with a patient population, and inter-rater reliability testing between four raters of varying experience without controlling moment. All raters were blinded.
Study Design: A series of trials to test intra-rater and inter-rater reliabilities. Results: Intra-rater reliability intraclass correlation coefficient was 0.98 and inter-rater reliability intraclass correlation coefficient (2,1) was 0.953 with a controlled moment. With uncontrolled moment, very high reliability for intra-tester was also achieved (intraclass correlation coefficient = 0.94 with knees extended and intraclass correlation coefficient = 0.95 with knees flexed). For the trial investigating test–retest reliability with actual patients, intraclass correlation coefficient of 0.99 was obtained. In the trial investigating four different raters with uncontrolled moment, intraclass correlation coefficient of 0.91 was achieved.
Conclusions: The new ankle dorsiflexion measurement device is a valid and reliable device for measuring ankle dorsiflexion in both healthy subjects and patients, with both controlled and uncontrolled moments, even by multiple raters of varying experience when the foot is dorsiflexed to its end of range of motion.
URI: https://www.um.edu.mt/library/oar/handle/123456789/104431
Appears in Collections:Scholarly Works - FacHScPod

Files in This Item:
File Description SizeFormat 
Validity_and_reliability_of_a_new_ankle_dorsiflexion_measurement_device(2013).pdf
  Restricted Access
450.79 kBAdobe PDFView/Open Request a copy


Items in OAR@UM are protected by copyright, with all rights reserved, unless otherwise indicated.